Products
4 Kelvin Closed Cycle Refrigerator Systems from Janis and SHI
Sumitomo (SHI) Compressor Units - 1.0-Watt & 1.5-Watt Compressors
Overview
The F-50 and CSA-71 are compatible with the 1-Watt (Model RDK-408D2), the 1.5-Watt (Model RDK-415D), the 6 K (Model RDK-408S), and the 0.5-Watt two-stage pulse tube cryocooler (Model RP-062B) cold head systems. The F-70 is compatible with the 1-Watt (Model RDK-408D2) and the 1.5-Watt (Model RDK-415D). All models are equipped with casters for ease of transport.
Standard Configuration
The F-50 and CSA-71 are compatible with the 1-Watt (Model RDK-408D2), the 1.5-Watt (Model RDK-415D), the 6 K (Model RDK-408S), and the 0.5-Watt two-stage pulse tube cryocooler (Model RP-062B) cold head systems. The F-70 is compatible with the 1-Watt (Model RDK-408D2) and the 1.5-Watt (Model RDK-415D). All models are equipped with casters for ease of transport.
Please click here to download a mechanical drawing for the F-50 compressor (188 KB).
The downloaded file will be in PDF format which is viewable and printable with Acrobat Reader, a free plug-in utility from Adobe. To download the free acrobat plug-in software click here.
Click here to contact Janis today!
 |
| Model F-50 Water-cooled Compressor |
 |
| Model F-70 Water-cooled Compressor |
 |
| Model CSA-71A Air-cooled Compressor |
Accessories |
| Input Power Cable | 5-Meter |
| Flexible Gas Lines | 6-Meter (RDK-408D, RDK-408S)
6-Meter with Ballast Tanks or
20-Meter (RDK-415D) |
| Cold Head Power Cable | 6-Meter |
Specifications
COMPRESSOR UNITS
| CSA-71A (UL/CE) |
| Cooling | Air |
| Ambient | 5 - 35 °C |
| Operating Range | 5% Max Capacity Loss from 28 - 35 °C |
| Electrical Power | AC 200V, 3 Phase |
| Electrical Power Consumption | Steady | Maximum |
| 50 Hz | 6.5 kW | 7.2 kW |
| 60 Hz | 7.5 kW | 8.3 kW |
| Weight | 150 kg |
| Dimension (mm) | 550D x 550L x 885H |
| Maintenance Interval | 20,000 hours |
| F-70L (UL/CE) | F-70H (UL/CE) | F-50L (UL/CE) | F-50H (UL-CE) |
| Cooling | Water | Water | Water | Water |
| Min. Flow Rate | 6-9 liters/minute | 6-9 liters/minute | 7 liters/minute | 7 liters/minute |
| Water Connection | 1/2" NPT Male | 1/2" NPT Male | 3/8" NPT Female | 3/8" NPT Female |
| Temperature Range | 5 - 25 °C | 5 - 25 °C | 4 - 28 °C | 4 - 28 °C |
| Electrical Power | AC 200V, 3 Phase | AC 380 - 415V, 3 Phase /
50 Hz
AC 480V, 3 Phase /
60 Hz | AC 200V, 3 Phase | AC 380, 400, 415V, 3 Phase / 50 Hz
AC 460 - 480V, 3 Phase /
60 Hz |
| Electrical Power Consumption | Steady | Maximum | Steady | Maximum | Steady | Maximum | Steady | Maximum |
| 50 Hz | 6.6-6.9 kW | 8.2-8.5 kW | 6.6-6.9 kW | 8.2-8.5 kW | 6.5 kW | 7.2 kW | 6.5 kW | 7.2 kW |
| 60 Hz | 7.5-7.8 kW | 8.7-9.0 kW | 7.5-7.8 kW | 8.7-9.0 kW | 7.5 kW | 8.3 kW | 7.5 kW | 8.3 kW |
| Weight | 100 kg | 100 kg | 120 kg | 120 kg |
| Dimension (mm) | 532H x 443W x 493D | 532H x 443W x 493D | 591H x 450W x 588D | 591H x 450W x 588D |
| Maintenance Interval | 30,000 hours | 30,000 hours | 30,000 hours | 30,000 hours |