Products

Liquid Helium (ST & STVP) and Liquid Nitrogen (ST)

Basic System Specifications - Continuous Flow Cryostats


Specification SuperTran
(Sample in Vacuum)
SuperTranVP
(Sample in Vapor)
Temperature Range <2-325 K (ST series: 100 - 400)
(500 K, >700 K optional)
1.5-325 K
(consult with Janis for other temperatures)
Initial Cooldown Time 15 minutes
(ST series: 100 - 400)
25 minutes
(ST-500)
15 minutes
(STVP series: 100-300)
30 minutes
(STVP-NMR)
Nominal Temperature Stability (with controller) 50 mK or less 50 mK or less
Orientation any position vertical for <4.5 K
System Weight (without transfer line) 10 lbs (4.6 kg) 15 lbs (6.8 kg)
Cryogen Consumption on Cooldown 0.4L LHe (325 - 4.2 K)
0.1L LN2 (325 - 77 K)
0.5L LHe (325 - 4.2 K)
Nominal Cryogen Consumption Rate 0.5 L/Hr LHe (5 K)
(ST series: 100-400)
1.1 L/Hr LHe (5 K) (ST-500)
0.1 L/Hr LN2 (100 K)
1.1 L/Hr LHe (5 K)
(STVP series: 100-300)
1.4 L/Hr LHe (5 K)
(STVP-NMR)
NOTE: Specifications do not include optical or experimental heat loads and assume vertical orientation.
NOTE: Consult Janis for high temperature or LN2 operations for SuperTran-VP.

Typical Cryogen Consumption of Optical SuperTran Cryostats

Click button to enlarge view of SuperTran and SuperTranVP liquid helium consumption rate, Janis Research, cryogenic equipment

 

 

Enlarge