Model HE-3-SOSV-S Cryostat
|
|
Description:
The model HE-3-SOSV-S is a
special optical cryostat. Optical windows with typical clear view
size of 0.5" are installed on the vacuum shroud, as well as both the
liquid nitrogen and liquid helium temperature radiation shields.
A special sample positioner
with rotary/linear motion manipulator is installed at the top of the
sample positioner so that the sample can be moved up and down, as well
as rotated. OFHC copper braids are employed to provide the
flexibilities of motion and rotation. A double optical sample
holder is installed at the bottom
of the sample positioner for reference
and
sample measurements.
This He-3 cryostat includes
the following components:
- Highly efficient liquid
nitrogen shielded helium dewar
- Liquid nitrogen temperature
radiation with optical windows
- Liquid helium temperature
radiation with optical windows
- 1 K radiation shield
- Charcoal sorption pumps
with heaters and thermometers
- 1 K pots with thermometers
- He-3 pots with heaters and
thermometers
- He-3 gas
- Sample probe with rotary/linear
motion manipulator
- Sample mount with heater
and thermometer
- Operation test
|
Options:
- Resistive wires
- Coaxial cables
- Automatic temperature
controller
- Helium level sensor and
read out
- Liquid helium transfer line
- 1 K pot pumping station
- High vacuum pumping station
|
Performance Specification with
two sets of 0.25" clear view quartz windows installed:
|
Sample Mount Base
Temperature:
|
less
than 350 mK |
| Holding
Time at Base Temperature:
|
24 hours minimum |
|
Operation Temperature Range:
|
0.35 K - 300 K |
|
Typical Sample Space:
|
1.5" |
|
Sample Z-motion:
|
+/- 0.5" |
|
Sample Rotation:
|
+/- 90 degree |
|
|


|
Table of Contents |
Pages |
Introduction |
1 |
Standard He-3 Cryostats |
HE-3-SSV Insert, Standard Sample Loading, Sample in Vacuum |
2 |
HE-3-SVSD Insert, Standard Sample Loading, Sample in Vacuum for He-4 Storage Dewar
 |
3 |
HE-3-SSXGAS Insert, Standard Sample Loading, Sample in Exchange Gas |
4 |
HE-3-XGHS Mini-Exchange-Gas Handling System |
5 |
HE-3-SOSV Cryostat, Standard Sample Loading, Optical, Sample in Vacuum |
6 |
HE-3-SOSV-S Cryostat, Standard Sample
Loading, Optical, Sample in Vacuum |
7 |
HE-3-SVC Cryostat, Standard Sample Loading, Sample in Vacuum, Compact for Detector Application |
8 |
HE-3-CCR-SSV Cryostat, Standard Sample Loading, Sample in Vacuum, Cryogen-free |
9 |
HE-3-SSV-CF Cryostat, Standard Sample Loading, Sample in Vacuum, Continuous Flow |
10 |
HE-3-TLSL Cryostat, Top Loading, Sample in He-3 Liquid |
11 |
HE-3-TLOSL Cryostat, Top Loading, Optical, Sample in He-3 Liquid |
12 |
HE-3-TLSV
Cryostat, Top Loading, Sample in Vacuum |
13 |
Gas Handling System for Top Loading He-3 Systems |
14 |
Special He-3 Cryostats |
HE-3-TLSL-NMR Cryostat, Top Loading, Sample in He-3 Liquid for NMR Experiments |
15 |
HE-3-TLSUHV-SPM Cryostat, Top Loading, Sample in UHV for Scanning Probe Microscope System |
16 |
HE-3-SSUHV-SPM/ESR Cryostat, Standard Sample Loading, Sample in UHV for Scanning Probe Microscope System and Electron Spin Resonance System |
17 |
HE-3-BLSUHV-STM Cryostat,
Bottom Loading, Sample in UHV for Scanning Tunneling Microscope System
|
18 |
JANIS RESEARCH COMPANY, INC.
2 Jewel Drive, P.O. Box 696
Wilmington, MA 01887-0696 USA
TEL: +1-978-657-8750 FAX: +1-978-658-0349
E-Mail: sales@janis.com
Policies
|
|