| Specifications |
CCS-100/202 |
CCS-100/204 |
|
Temperature range |
11 K - 325 K |
9 K - 325 K
(optional 6.5 K - 325 K
with CCS-100/204N) |
Refrigeration
capacity
(typical without attachments) |
60
Hz |
0.5 W @
10 K |
2.0 W @
10 K |
| 2.5 W @
20 K |
9.0 W @
20 K |
| 10.0 W @
77 K |
17.0 W @
77 K |
| |
| 50
Hz |
2.0 W @
20 K |
7.5 W @
20 K |
| 8.3 W
@ 77 K |
14.0 W @
77 K |
| Cooldown
time (typical) |
60
Hz |
50
minutes to 20 K |
30
minutes to 20 K |
| 50
Hz |
60
minutes to 20 K |
35
minutes to 20 K |
| System
weight |
Compressor |
75 kg
(165 lbs) |
75 kg
(165 lbs) |
| Expander |
6.8 kg
(15 lbs) |
7.7 kg
(17 lbs) |
| Power
supply
|
Volts
(VAC±5%) / Phase (Ø) /
Frequency (Hz) |
208/230
/ 1 / 60
200 / 1 / 50
230/240 / 1 / 50 |
| Power
requirements |
3.0 kW, 15.5 amperes full load at 208
V~, 60 Hz and 2.6 kW, 13 amperes full load at 200 V~, 50 Hz.
58 amperes starting current,
68 amperes locked-rotor current. |
| Water-cooled
compressor requirements:
|
Flow |
2.7 lpm (0.7 gpm) |
| Temperature |
4-27 °C |
|
Chiller requirement |
3.2
kW (11,000
BTU/hr) |
|
Pressure |
30-100 psig |


Please
click here for the standard features and available options
of Models CCS-100/202 & CCS-100/204.
|
Table of Contents | | Pages |
| Introduction
| | 1,
2 |
| CCS-150
& CCS-450 Standard Optical Systems
| | 3, 4 |
| CCS-100/202 & CCS-100/204 Standard Optical Systems | | 5, 6 |
| CCS-350,
CCS-350S & CCS-350R Compact Optical Systems
| | 7, 8 |
| CCS-250,
CCS-350T & CCS-350ST Non-Optical Systems
| | 9, 10 |
| CCS-550,
CCS-550T, CCS-950, CCS-950T Top Loading Exchange Gas Systems
| | 11, 12 |
| CCS-750
Vibrating Sample Magnetometry Systems
| | 13 |
| CCS-850
Mössbauer Spectroscopy Systems | | 14,
15 |
| Other
Systems (including CS-1020)
| |
16 |
| Model 8200
Compressor Specifications
| |
17 |
| Options and Accessories
| |
18 |
|
|