The Janis ST-500-LN
continuous flow cryostat has been specially designed for use in
microscopy, imaging, and high spatial resolution photoluminescence.
The combination of low thermal-expansion support structure and
internal vibration isolation results in ultra-low vibration and drift
levels. The cryostat geometry offers a short working distance
(for use with high magnification optics), and permits mounting on
common microscope stages and translators. This low-profile
cryostat is only 1.52" tall (<39 mm).
The ST-500-LN
system includes a flexible high-efficiency variable flow cryogen
transfer line, which delivers liquid helium or liquid
nitrogen to a heat exchanger and sample mount.
The ST-500-LN is portable, easily mounted in a variety of
configurations, and can be used both in transmission and reflection
geometries.
Contact us to find out
how easy it is to afford one of these systems!
Please click here to view the ST-500-LN
Microscopy Cryostat
Mechanical Drawing.
Options include:
-
A variety
of window materials and sizes
-
Interchangeable
sample mounts for different sample thicknesses
-
Electrical
and RF feedthroughs
-
A complete
system solution including temperature controllers, pumping
systems and storage dewars.
LHe Specifications
|
Temperature range (continuous)
w/LHe
|
<10 - 325 K
(standard)
|
| |
<10 - 450 K (optional)* |
| Initial cooldown time |
10 K (LHe):
25 minutes |
| Cryogen usage on cooldown |
1
L
LHe |
| Temperature
stability |
<50 mK |
| (with temperature controller) |
|
| Weight |
6 lbs.
(2.7 kg) |
| (not including transfer line) |
|
| Nominal cryogen usage |
LHe @ 10 K:
~2 L/hr |
| |
LHe @ 20 K:
~1.5 L/hr |
| Nominal vibrational amplitude |
±25 nm expected |
| Positional
drift |
±50 nm in 5 minutes
expected |
|
|
±150 nm
in 30 minutes expected
|
|
*
With optional equipment, consult with
Janis for details.
|
Cryostat
overall height (thickness)
|
1.52" (39 mm)
|
|
Temperature range (continuous) w/LN2 |
77 -
325 K (standard) |
| |
77 -
450 K (optional)* |
|
Initial cooldown time |
77
K: 15 minutes |
|
Cryogen usage on cooldown |
<0.3
L
LN2 |
|
Temperature stability |
<50 mK |
| (with
temperature controller) |
|
|
Weight |
6 lbs.
(2.7 kg) |
| (not
including transfer line) |
|
|
Nominal cryogen usage |
~55 mL/hr |
|
*
With optional equipment, consult with
Janis for details.
|