Janis Research, cryogenics, cryostats

Basic System Specifications

Specification SuperTran
(Sample in Vacuum)
SuperTran VP
(Sample in Vapor)
Temperature Range <2-325K
(500 K, >700 K optional)
1.5-325K
(consult with Janis for
other temperatures)
Initial Cooldown Time 15 minutes
(ST series: 100-400)
25 minutes
(ST-500, ST-500-LN)
15 minutes
(STVP series: 100-400)
30 minutes (STVP-NMR)
Nominal Temperature
Stability (with controller)
50 mK or less 50 mK or less
Orientation any position vertical for < 4.5 K
System Weight
(without transfer line)
10 lbs (4.6 kg) 15 lbs (6.8 kg)
Cryogen Consumption
on Cooldown
0.4L   LHe (325 - 4.2 K)
0.1L   LN2 (325 - 77 K)
0.5L   LHe (325 - 4.2 K)
Nominal Cryogen
Consumption Rate
0.5 L/Hr   LHe (5 K)
(ST series: 100-400)
1.1 L/Hr   LHe (5 K) (ST-500)
2.2 L/Hr   LHe (10 K) (ST-500-LN)
0.1 L/Hr   LN2 (100 K)
1.1 L/Hr   LHe (5 K)
(STVP series: 100-400)
1.4 L/Hr   LHe (5 K)
(STVP-NMR)
NOTE:   Specifications do not include optical or experimental heat loads and
             assume vertical orientation.
NOTE:   Consult Janis for high temperature or LN2 operations for SuperTran-VP.

Typical Cryogen Consumption of Optical SuperTran Cryostats

typical cryogen consumption of optical SuperTran continuous flow cryostat by Janis Research, cryogenics


Table of Contents

Page
Introduction 1, 2
ST-100 Cryostat Systems 3, 4
ST-100-FTIR Cryostat Systems 5
ST-170 ESR Systems 6
ST-200 Tubular Systems 7
ST-200-NMR Tubular Systems 8
ST-300/300T Compact Systems 9, 10
ST-300-MS Compact Systems for Microscopy   11, 12
ST-400 Ultra High Vacuum Systems 13
ST-500 Microscopy Cryostat

14, 15

ST-500-LN Microscopy Cryostat

16, 17

Basic System Specifications

18

Options & Accessories

19

Next Page Janis Products Contact Janis Research Back to Janis Research Home Page