| Specification |
SuperTran (Sample in Vacuum) |
SuperTran VP (Sample in Vapor) |
| Temperature Range |
<2-325K (500 K, >700 K optional) |
1.5-325K (consult with Janis for other temperatures) |
| Initial Cooldown Time |
15 minutes
(ST series: 100-400)
25 minutes
(ST-500, ST-500-LN) |
15 minutes
(STVP series: 100-400)
30 minutes (STVP-NMR) |
Nominal Temperature Stability (with controller) |
50 mK or less |
50 mK or less |
| Orientation |
any position |
vertical for < 4.5 K |
System Weight (without transfer line) |
10 lbs (4.6 kg) |
15 lbs (6.8 kg) |
Cryogen Consumption on Cooldown |
0.4L LHe (325 - 4.2 K) 0.1L LN2 (325 - 77 K) |
0.5L LHe (325 - 4.2 K) |
Nominal Cryogen Consumption Rate |
0.5 L/Hr LHe (5 K)
(ST series: 100-400)
1.1 L/Hr LHe (5 K) (ST-500)
2.2 L/Hr LHe (10 K) (ST-500-LN) 0.1 L/Hr LN2 (100 K) |
1.1 L/Hr LHe (5 K)
(STVP series: 100-400)
1.4 L/Hr LHe (5 K)
(STVP-NMR) |
NOTE: Specifications do not include optical or experimental heat loads
and
assume
vertical orientation.
NOTE:
Consult Janis for high temperature or LN2 operations for SuperTran-VP.
|