The Janis VNF-100 cryostat is a liquid
nitrogen variable temperature cryostat with the sample located in flowing vapor.
Ideal for experiments with samples that are difficult to thermally anchor such
as liquid or powder samples, the VNF-100 is a low-cost alternative to the
SuperTran-VP or SuperVariTemp dewars when temperatures below 65 K are not
required. The VNF-100 features a top-loading sample chamber for rapid sample
exchange, and four-way, f = 1.0, optical access to the sample chamber.
Specifications
| LN2 capacity (nominal) |
1.2
 |
| Working time (typical) |
approximately 6 - 8 hours |
| Initial cooldown time (from room
temperature) |
approximately
30 minutes |
| Sample exchange time (typical) |
less than 5 minutes |
| Operating temperature range |
65 - 325 K (<50 K in N2 “ice”) |
| Temperature stability (with
temperature controller) |
50 mK or better |
|
Standard Features:
- Quick refill capability with
included funnel assembly.
- High quality bellows-sealed
evacuation valve and a built-in cryopump for maintaining high vacuum.
- Safety pressure relief valves
protecting all independent spaces.
- 1.25" diameter sample tube with
customer-specified optical access (standard system offers four quartz window
pairs with
ƒ = 1.0 optical access to the sample space). - Top loading sample chamber for rapid
sample exchange.
- Variable cooling system which places
the sample in flowing N2 vapor (ideal for low thermal conduction
samples), providing excellent cooling power and temperature control.
- Light-weight sample positioner/mount
assembly, with multi-pin electrical feedthrough access to the window region,
offering rotation and translation around the cryostat axis.
- Two silicon diodes installed on the
vaporizer (heat exchanger) and sample mount, for controlling and monitoring
the system temperature.
Optional Equipment:
- Exhaust flow gauge for more precise
control.
- Special window materials and
geometries (IR, UV, and x-ray).
- Special sample holders including
four-point resistivity measurement and liquid cuvettes.
Please go to the next page to view the Model VNF-100
Mechanical Drawing.
Contact Janis
today for details and pricing.
 |
 |
Model VNF-100T
|
Special 475 K High Temperature
Non-optical VNF-100T
|
 |
The
photo on the left shows a cuvette sample holder for installation on an optical
VNF-100. It is compatible with all 12.5mm wide x 12.5mm deep cuvettes such
as the Starna models 1-G-10, 1.7-G-10, 1-SOG-10, 1-Q-10, and 1-I-10.
(Click on the photo for a larger view.) |
|