General Measurement
Number of inputs: 3
Update rate: Up to 4 rdg/s on display; up to 18 rdg/s with IEEE-488 interface
Measurement modes: DC, RMS, peak
Probe compatibility: Standard, multi-axis, and custom probes
Probe features: Linearity Correction, Temperature Correction, Auto Probe Zero
Measurement features: Autorange, Max Hold, Relative Mode, Filter, Vector Magnitude, Differential Reading
Probe connector: 15-pin D style
DC measurement
Probe type ranges | 5¾-digits with filter | 4¾-digits without filter |
HST probe |
300 kG | 0.001 kG | 0.01 kG |
30 kG | 0.0001 kG | 0.001 kG |
3 kG | 0.00001 kG | 0.0001 kG |
300 G | 0.001 G | 0.01 G |
HSE probe |
30 kG | 0.0001 kG | 0.001 kG |
3 kG | 0.00001 kG | 0.0001 kG |
300 G | 0.001 G | 0.01 G |
30 G | 0.0001 G | 0.001 G |
UHS probe (discontinued) |
30 G | 0.0001 G | 0.001 G |
3 G | 0.00001 G | 0.0001 G |
300 mG | 0.001 mG | 0.01 mG |
DC accuracy: ±0.10% of reading ±0.005% of range
DC temperature coefficient: ±0.05% of reading ±0.003% of range per ºC
AC RMS & peak measurement
AC display resolution: 4¾ digits
Probe type ranges | RMS resolution | Peak resolution |
HST probe |
300 kG | 0.01 kG | 0.01 kG |
30 kG | 0.001 kG | 0.001 kG |
3 kG | 0.0001 kG | 0.0001 kG |
300 G | 0.01 G | × |
HSE probe |
30 kG | 0.001 kG | 0.001 kG |
3 kG | 0.0001 kG | 0.0001 kG |
300 G | 0.01 G | 0.01 G |
30 G | 0.001 G | × |
UHS probe (discontinued) |
30 G | 0.001 G | 0.001 G |
3 G | 0.0001 G | 0.0001 G |
300 mG | 0.01 mG | × |
AC frequency range: 10 Hz to 400 Hz
AC RMS accuracy: ±2% of reading (50 Hz to 60 Hz)
AC RMS frequency response: 0 to -3.5% of reading (10 Hz to 400 Hz)
(All AC RMS specifications for sinusoidal input >1% of range)
AC peak accuracy: ±5% typical
AC peak speed: 5 ms for single peak
Front panel
Display type: 4-line × 20-character, vacuum fluorescent
Display resolution: Up to ±5¾ digits
Display update rate: 4 rdg/s with vector off, 3 rdg/s with vector on
Display units: Gauss (G), Telsa (T)
Units multipliers: µ, m k
Annunciators: RMS: AC input signal, DC: DC input signal, MAX: Max Hold value, s: Relative reading, R: Remote operation, ?:Alarm on
Keypad: 25 full travel keys
Front panel features: Intuitive operation, display prompts, front panel lockout, brightness control
Interfaces
RS-232C capabilities
Baud: 300, 1200, 9600
Connector: RJ-11 configuration
Update rate: Up to 14 rdg/s at 9600 baud
IEEE-488 capabilities
Complies with IEEE-488.2 SH1, AH1, SR1, RL1, PP0, DC1, DT0, C0, E1
Update rate: 18 rdg/s with vector off, 14 rdg/s with vector on
Alarm
Settings: High and low setpoint, Inside/Outside, Audible
Actuators: Display annunciator, beeper
Monitor analog output (3)
Configuration: Real time analog voltage output
Scale: ±3 V = ±FS on selected range
Frequency response: DC to 400 Hz
Accuracy: Probe dependent
Minimum load resistance: 1 kΩ (short circuit protected)
Connector: BNC
Corrected analog output (1)
Configuration: Voltage output generated by DAC
Range: ±3 V; ±10 V for the Model 460-10
Scale: User-defined
Resolution: 0.366 mV of ±3 V
Update rate: Same as field measurement
Accuracy: ±0.1% full scale in addition to measurement error
Minimum load resistance: 1 kΩ (short circuit protected)
Connector: BNC
General
Ambient temperature: 15 to 35 ºC at rated accuracy. 5 to 40 ºC with reduced accuracy
Power requirement: 100, 120, 220, 240 VAC (+5%, -10%), 50 or 60 Hz, 40 VA
Size: 432 mm W × 89 mm H × 368 mm D (17 in × 3.5 in × 14.5 in), full rack
Weight: 7.5 kg (16.5 lb)